Short answer
An autoclave is a device that sterilises surgical instruments and heat-resistant materials with saturated steam under pressure. In the sealed chamber the temperature of the steam rises above 100 °C together with the pressure; this hot, moist steam destroys bacteria, viruses, fungi and the most resistant form, bacterial spores. The two cycles most used in clinics are the 121 °C and 134 °C cycles; which cycle is used for which load is set by the manufacturer of the device and by the clinic's sterilisation procedure.
The autoclave alone is not enough: the instrument is first cleaned, dried, packed in a sterilisation reel pouch or sterilisation paper or placed in a sterilising drum, and after the cycle it is stored dry and closed until the pack is opened. The Demir Med catalogue lists the Autoclave 22 L, the Sterilisation Reel 30 cm, the Sterilisation Paper 30 cm and the Sterilising Drum; all are sold by quotation.
What is an autoclave and how does it work?
The shortest answer to what an autoclave is: a pressurised steam steriliser. It relies on the same physics as a pressure cooker. In an open pot water boils at 100 °C and gets no hotter; in a sealed, pressurised chamber the temperature of the steam rises. The killing effect of an autoclave does not come from the pressure but from the saturated steam at high temperature that the pressure makes possible. As the steam condenses on the instrument it transfers its heat quickly and denatures the proteins of microorganisms irreversibly.
An autoclave cycle has three stages:
| Stage | What happens | Why it matters |
|---|---|---|
| Air removal and heating | The air in the chamber is removed and steam fills the chamber | Steam cannot reach a place where air remains; that spot is not sterilised |
| Sterilisation (holding) | The chamber is held at the selected temperature for the cycle time | The whole time is needed for the point of the load that heats last |
| Exhaust and drying | The steam is released and the load is dried | A pack that comes out wet lets microorganisms through and is not sterile |
How the air is removed defines the class of the autoclave, which is covered below.
What is sterilisation? Sterilisation versus disinfection
Sterilisation is the destruction of all microorganisms on an item, including bacterial spores. Disinfection is the destruction of most disease-causing microorganisms on inanimate surfaces; it does not kill all spores. The difference between sterilisation and disinfection is therefore not one of degree but of result: no living microorganism remains on a sterile instrument, while on a disinfected surface the number has been brought down to a safe level.
Which process applies to which item depends on where the item touches the body:
| Item group | What it touches | Process required | Example |
|---|---|---|---|
| Critical | Sterile tissue, the vascular system | Sterilisation | Surgical forceps, needle holder, scissors, scalpel handle |
| Semi-critical | Mucous membranes, non-intact skin | Sterilisation or high-level disinfection | Some heat-sensitive examination instruments |
| Non-critical | Intact skin | Low or intermediate-level disinfection | Blood pressure cuff, examination couch surface |
This classification is a general framework; the clinic's infection control procedure and the manufacturer's instructions for the instrument always come first.
Sterilisation methods
The autoclave is the most common method, but not the only one. The sterilisation methods compare as follows:
| Method | How it works | Suitable load | Limit |
|---|---|---|---|
| Steam (autoclave) | Saturated steam under pressure | Stainless steel instruments, textiles, autoclavable materials | Not suitable for items that cannot take heat and moisture |
| Dry heat | Hot dry air at high temperature for a long time | Glass, metal, oils and powders | Long exposure; unsuitable for most textiles, plastics and packed loads |
| Low-temperature methods (ethylene oxide, hydrogen peroxide plasma) | Gas or plasma at low temperature | Heat-sensitive devices and materials | Needs dedicated equipment and infrastructure; usually hospital scale |
| Liquid chemical (immersion) | The instrument stays in the solution for the time on the label | Heat-sensitive instruments that cannot be autoclaved | Gives no packed sterility; the instrument is prepared just before use |
For every heat-resistant instrument, steam is the first choice. Immersion is used for instruments that cannot go into an autoclave; the Cidex 2% glutaraldehyde instrument disinfectant in our catalogue is supplied for this purpose: it is prepared with its activator, the activated solution can be used for 14 days in a closed container, and the instrument is kept fully immersed for the time stated on the label.
Autoclave temperature and time
The answer to what temperature an autoclave runs at depends on the cycle. Two temperatures have become standard in bench-top autoclaves:
| Cycle | Temperature | Approximate pressure | Typical holding time | Load it is generally used for |
|---|---|---|---|---|
| 121 °C cycle | 121 °C | About 1 bar (above atmospheric) | 15–20 minutes | More heat-sensitive autoclavable materials, some plastic and rubber parts |
| 134 °C cycle | 134 °C | About 2 bar (above atmospheric) | 3–5 minutes | Stainless steel surgical instruments, textiles |
These are general reference values and show the holding stage only; with heating, air removal and drying a full cycle usually takes 30–60 minutes. What counts is the cycle validated by the manufacturer of the device: nobody leaves the cycle to shorten the time or change the temperature. The instrument sets a limit too; whether an instrument withstands 134 °C is stated in its manufacturer's instructions for use.

Autoclave cycles
121 °C · 134 °C
Autoclave classes: N, S and B
Small steam sterilisers are divided into three classes by how they remove the air and which loads they can sterilise:
| Class | Air removal | Load it can sterilise | Who it is for |
|---|---|---|---|
| Class N | By gravity (steam displaces the air) | Unwrapped, solid instruments only | Simple instruments that are not packed and are used straight away |
| Class S | The method specified by the manufacturer | The loads defined by the manufacturer | Devices chosen for specific instrument types |
| Class B | Pre-vacuum (the air is pumped out) | All loads, including wrapped, hollow and porous loads | Clinics that sterilise packed instruments, textiles and hollow instruments |
A class B autoclave is preferred in most clinics because instruments are sterilised packed and stored packed. In hollow instruments (cannulated or tube-shaped parts, for example) and in porous loads such as swabs inside a drum, the air is fully removed only by vacuum. For the Autoclave 22 L in our catalogue the class, the cycles and the drying function are stated in writing in the quotation; if you will sterilise wrapped, hollow and porous loads, always ask for the class.
How to sterilise with an autoclave, step by step
Autoclave sterilisation is a process, not a device function. If one step is skipped, the device finishing its cycle does not mean the instrument is sterile.
- Pre-cleaning: The used instrument is freed of gross soil before it dries. Dried blood and tissue are harder to remove and steam cannot get under the soil.
- Cleaning: Instruments are cleaned with their joints open, brushed with a suitable cleaning solution or in an ultrasonic cleaner. An instrument that has not been cleaned cannot be sterilised.
- Rinsing and drying: Detergent residue is rinsed off and the instrument is dried completely. An instrument packed wet stains and corrodes.
- Inspection: Jaw closure, ratchet, joint and cutting edges are checked; a damaged instrument is taken out of the set.
- Packaging: The instrument goes into a pouch cut from the reel, into sterilisation paper or into a drum; ratcheted instruments are left open.
- Loading: Packs are not squeezed on top of each other; a gap is left for steam to pass between them. Reel pouches are placed with the paper side facing the same way.
- Cycle: The cycle that suits the load is selected and runs to the end without interruption.
- Drying and unloading: Packs must come out dry. A wet, torn or opened pack is not accepted as sterile and is processed again.
- Record and storage: The cycle, the date and the indicator result are recorded; packs are stored in a dry, closed cupboard away from dust.

The stainless steel instrument tray and kidney dish are reusable items that are sterilised in the autoclave; they are used to prepare instruments and to carry used instruments to the sterilisation area.

Sterilisation steps
Clean · pack · sterilise · store
Packaging: sterilisation reel, sterilisation paper and the drum
Sterilisation packaging has two jobs: to let steam in during the cycle and to keep the instrument sterile until the pack is opened.
| Packaging | How it is used | For which load |
|---|---|---|
| Sterilisation reel | Cut to the length needed and sealed at both ends to make a pouch | A single instrument or small groups: forceps, needle holder, scissors |
| Sterilisation paper | The instrument set or tray is wrapped in the paper | Instrument sets, trays, large packs |
| Drum | The material is placed inside the lidded container | Swabs, gauze, cotton, drapes and instruments |
A sterilisation reel is packaging with paper on one side and transparent film on the other; the instrument is visible inside and steam passes through the paper side. The width of the reel is chosen by the length of the instrument to be packed: the instrument must not sit tight in the pouch, and a gap must remain between the seal and the instrument. In our catalogue the Sterilisation Reel 30 cm and the Sterilisation Paper 30 cm are separate products; the length, the material, the indicator and the suitable sterilisation method are stated on the label and in the quotation.

A sterilising drum is the lidded, cylindrical container in which gauze, cotton, drapes and instruments are sterilised and kept until use. In classic drums the holes in the body are opened during sterilisation and closed with a sliding band when the cycle is over. Every time a drum is opened its contents meet room air; clinics therefore define in their own procedures how long a drum may be used after it has been opened. For the Sterilising Drum in our catalogue the diameter, height and material are stated in the quotation.
Autoclave tape and indicators: how is sterilisation checked?
You cannot see whether a pack has been through the autoclave; indicators are used for that. Three kinds of check complement each other:
| Check | What it shows | What it does not show |
|---|---|---|
| Physical (display, printout) | That the cycle reached its temperature, pressure and time | That steam reached the inside of the pack |
| Chemical (autoclave tape, indicator printed on the pouch, indicator inside the pack) | That the pack has been processed; with internal indicators, that the conditions reached the pack | It does not prove on its own that microorganisms were killed |
| Biological (spore test) | That resistant bacterial spores were killed | It does not verify each pack; it is run at set intervals |
Autoclave tape is used to close a pack and to show that it has been processed; the stripes on it change colour with steam. The colour change shows that the pack has been in the autoclave; it does not prove on its own that the instrument inside is sterile. In pre-vacuum autoclaves a separate test (Bowie-Dick) shows that air removal works. Which test is run and how often is set by the clinic's sterilisation procedure and the applicable regulations; the results are recorded.
What goes into an autoclave and what does not?

| Sterilised in the autoclave | Not for the autoclave |
|---|---|
| Stainless steel surgical instruments: needle holder, forceps, scissors, scalpel handle | Single-use products: syringes, needles, scalpel blades, single-use punches |
| Stainless steel trays, kidney dishes and bowls | Plastic and electronic parts that cannot take heat |
| Textiles and drapes defined as autoclavable | Oils and powders (steam does not penetrate them) |
| Hand instruments and parts the manufacturer declares autoclavable | Motors, cables and adaptors for which the manufacturer forbids autoclaving |
Two rules do not change. First, a single-use product is not re-sterilised: a product with the single-use symbol on its packaging is made for one patient and is discarded after use. Second, the manufacturer's instructions govern every instrument: for instruments used in hair transplantation such as the micromotor handpiece, the DHI pen or the blade handle, the method and the temperature of sterilisation are written in the manufacturer's manual. For the instruments used in a hair transplant see the guides to the instruments used in hair transplantation and hair transplant forceps types.
Common mistakes in using an autoclave
- Putting a dirty instrument in the autoclave: Dried blood and tissue block the steam; the autoclave does not replace cleaning.
- Overloading the chamber: In a tight load steam cannot pass between the packs and drying is poor too.
- Sterilising a ratcheted instrument closed: Steam does not reach inside a closed joint; needle holders, clamps and scissors are packed open.
- Putting a wet pack on the shelf: Wet packaging lets microorganisms through; the pack is processed again.
- Using the wrong water: Autoclaves run on water of the quality the manufacturer describes (usually distilled or demineralised); tap water leaves scale and stains and damages the device.
- Relying on autoclave tape alone: The tape shows that the pack has been processed; sterility is checked by following physical, chemical and biological indicators together.
- Postponing maintenance: The door gasket, the chamber and the filter are cleaned at the intervals the manufacturer states; maintenance and validation records are kept.
Checklist for buying an autoclave and sterilisation supplies
- Does the chamber volume match your daily instrument load? The device in our catalogue has a 22-litre chamber.
- Is the class of the autoclave (B, S or N) stated in the quotation? Ask for the class if you sterilise wrapped, hollow and porous loads.
- Are the cycles, the drying function and the cycle times stated?
- Are the trays and accessories listed item by item in the quotation?
- Are the electrical requirements and the installation place suitable?
- Have you asked for the device's documents? CE and ÜTS registration details are shared once verified.
- Are the user and maintenance manual, service and spare parts terms clear?
- Was the width of the reel and paper chosen for your longest instrument?
- Are indicators, autoclave tape and test materials on the order list?
- Is a high-level disinfectant planned for instruments that cannot be autoclaved?
- Are a container for carrying used instruments, an instrument tray and a sharps container on the list?
For planning the sterilisation area when setting up a clinic see the sterilisation devices page, for choosing instruments the guides to surgical instrument names and needle holder types, and for sterile consumables the sterile medical consumables guide.
Autoclave and sterilisation supplies from Demir Med
Demir Med supplies clinics and hair transplant centres with the products of the sterilisation chain by quotation: the Autoclave 22 L is in the Medical Equipment category; the sterilisation reel, the sterilisation paper and the instrument disinfectant are in Disinfection and Hygiene; the sterilising drum, trays and dishes are in Surgical Instruments. Prices are not published on the site. Send your list through the quotation page; a written quotation is sent within 1 business day.